Medical Grade Plug-In Line Isolation Transformers

Medical Grade Plug-In Line Isolation Transformers 

178, 240, 289, 298 "MG" Series

*Not intended to be used to supply power to life support appliances
Medical Grade Plug-In Line Isolation Transformers

Features:

  • cULus Listed to UL 60601-1 (1st Edition) and CAN/CSA-C22.2 No. 601.1-M90 (R2005).
  • Each unit provides complete line isolation and noise filtering plus utilizes a Faraday shield to reduce leakage currents below UL60601-1 requirements (under 100 microamps).
  • Reduced noise coupling due to floating AC output.
  • Intended for non-patient connected indoor applications protecting sensitive electronic equipment. 
  • 3 different VA Capacities available - -  500VA, 1000VA, and 1800VA.  All rated 50/60Hz
  • Vented 14 gauge steel housing with white powder paint
  • Hospital Grade 3-wire grounded Input Plug (see chart).
  • 2 or 3 Duplex Hospital Grade Output Receptacles (see chart).
  • On-Off Switch / Circuit Breaker combo protected, and built in automatic Thermal Cut Out for added protection. 

4 different electrical configurations:

  • 120V-120V straight isolation ("178" series)
  • 240V-240V straight isolation ("240" series)
  • 240V-120V step down isolation ("289" series)
  • 120V-240V step up isolation ("298" series)

Part Details

Straight Isolation 178 Series

Part Number VA Input Voltage Input Current Output Voltage Output Current Leakage Current Full Load Efficiency Input Plug Cord Style / Length Receptacle Output # of Duplex
178MG500 500 120 VAC 4.33 A 120 VAC 4.17 < 100uA 93% (min) 5-15P 14/3 SJTW /6ft. Duplex 5-15R 2
178MG1000 1000 120 VAC 8.67 A 120 VAC 8.33 A < 100 uA 95% (min) 5-15P 14/3 SJTW /6ft Duplex 5-15R 2
178MG1800 1800 120 VAC 15.63 A 120 VAC 15.00 < 100 uA 96% (min) 5-20P 12/3 SJTW /6ft Duplex 5-15R 3

Straight Isolation 240 Series

Part Number VA Input Voltage Input Current Output Voltage Output Current Leakage Current Full Load Efficiency Input Plug Cord Style / Length Receptacle Output # of Duplex
240MG500 500 240 VAC 2.17 A 240 VAC 2.08 < 100uA 93% (min) 5-15P 14/3 SJTW /6ft. Duplex 6-15R 2
240MG1000 1000 240 VAC 4.33 A 240 VAC 4.17 A < 100 uA 95% (min) 5-15P 14/3 SJTW /6ft. Duplex 6-15R 2
240MG1800 1800 240 VAC 7.81 A 240 VAC 7.50 A < 100 uA 96% (min) 5-20P 14/3 SJTW /3ft Duplex 6-15R 3

Step-Down Isolation 289 Series

Part Number VA Input Voltage Input Current Output Voltage Output Current Leakage Current Full Load Efficiency Input Plug Cord Style / Length Receptacle Output # of Duplex
289MG500 500 240 VAC 2.17 A 120 VAC 4.17 < 100uA 93% (min) 6-15P 14/3 SJTW /6ft. Duplex 5-15R 2
289MG1000 1000 240 VAC 4.33 A 120 VAC 8.33 A < 100 uA 95% (min) 6-15P 14/3 SJTW /6ft. Duplex 5-15R 2
289MG1800 1800 240 VAC 7.81 A 120 VAC 15.00 < 100 uA 96% (min) 6-15P 14/3 SJTW /6ft Duplex 5-15R 3

Step-Up Isolation 298 Series

Part Number VA Input Voltage Input Current Output Voltage Output Current Leakage Current Full Load Efficiency Input Plug Cord Style / Length Receptacle Output #of Duplex
298MG500 500 120 VAC 4.33 A 240 VAC 2.08 < 100uA 93% (min) 5-15P 14/3 SJTW /6ft. Duplex 6-15R 2
298MG1000 1000 120 VAC 8.67 A 240 VAC 4.17 A < 100 uA 95% (min) 5-15P 14/3 SJTW /6ft. Duplex 6-15R 2
298MG1800 1800 120 VAC 15.63 A 240 VAC 7.50 < 100 uA 96% (min) 5-20P 12/3 SJTW /6ft Duplex 6-15R 3